FDA Recall Terminated

Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.

Recall: Z-2116-2017 · Initiated February 15, 2012

Recall

Recall Number
Z-2116-2017
Event Number
76985
Firm
Medtest Holdings, Inc.
FEI Number
1827821
Product Code
CIT
Status
Terminated
Root Cause
Error in labeling
Initiated
February 15, 2012
Posted
May 18, 2017
Terminated
June 2, 2017
Address
5449 Research Dr, Canton, MI, 48188-2261

Description

Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.

Reason

A Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the Performance section linearity is listed as 0-600 U/L, which is not consistent with the 510k submission package insert.

Action

Technical Bulletin Notification letter was sent 02/15/12 by email to all customers and distributors. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-757-5313.

Distribution

US Nationwide Distribution

Quantity

119.168L across All lots