FDA Recall Terminated

ACUSON Sequoia C512/512SS Adv PE 10.0 System, model number 10041127. Distributed by Siemens Medical Solutions USA, Inc., Business Unit Ultrasound, Mountain View, CA. The ACUSON Sequoia is a general purpose Diagnostic Ultrasound system. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in: B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, Harmonic Imaging or 3D imaging on a FPD Display. The ACUSON Sequoia is intended for the following applications: Abdominal, General Imaging and Cardiolgoy for Fetal, Abdominal, Intraoperative (abdominal and neurological), Pediatrics, Small Organs (breat, testes, thyroid and penis), Neonatal/Adult Cephalic, Cardiac (adult, pediatric and neonatal), Transesophageal, Transrectal, Transvaginal, Peripheral Vessels and Musculo-sceletal (superficial and conventional) applications. The system also provides for the measurement of anatomical structures that provide information that is used for clinical diagnosis purposes.

Recall: Z-2115-2011 · Initiated May 15, 2009

Recall

Recall Number
Z-2115-2011
Event Number
53013
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
2936884
Product Code
IYN
Status
Terminated
Root Cause
Release of Material/Component prior to receiving test results
Initiated
May 15, 2009
Posted
April 29, 2011
Terminated
April 29, 2011
Address
1230 Shorebird Way P.O. Box 7393, Mountain View, CA, 94043

Description

ACUSON Sequoia C512/512SS Adv PE 10.0 System, model number 10041127. Distributed by Siemens Medical Solutions USA, Inc., Business Unit Ultrasound, Mountain View, CA. The ACUSON Sequoia is a general purpose Diagnostic Ultrasound system. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in: B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, Harmonic Imaging or 3D imaging on a FPD Display. The ACUSON Sequoia is intended for the following applications: Abdominal, General Imaging and Cardiolgoy for Fetal, Abdominal, Intraoperative (abdominal and neurological), Pediatrics, Small Organs (breat, testes, thyroid and penis), Neonatal/Adult Cephalic, Cardiac (adult, pediatric and neonatal), Transesophageal, Transrectal, Transvaginal, Peripheral Vessels and Musculo-sceletal (superficial and conventional) applications. The system also provides for the measurement of anatomical structures that provide information that is used for clinical diagnosis purposes.

Reason

Shock Hazard. The manufacturer failed to set the system power switch to "ON" during the line level leakage and high pot current draw testing.

Action

Siemens Medical Solutions USA, Inc. sent a Customer Safety Advisory Notification letter out on May 15, 2009. Service engineers were to visit each customer to conduct necessary testing. For questions regarding this recall call 650-694-5993.

Distribution

Worldwide Distribution - USA including CA, IL, MD, NC, OH, and UT and the countries of Australia, China, Denmark, Spain, South Korea, Norway, Poland, Russia and Sweden.

Quantity

4 units