FDA Recall Terminated

Doctors Review System, OB-Windows/OB-View software release numbers 4.1 through 4.1.3, Digisonics, Inc., 3701 Kirby Drive, Houston, TX 77098 OB-Windows provides professional reporting and analysis of fetal growth throughout the gestational period, which can help evaluate clinical situations. OB-View provides image management for the user to capture, view, read, and edit images.

Recall: Z-2113-2011 · Initiated February 9, 2007

Recall

Recall Number
Z-2113-2011
Event Number
58367
Firm
Digisonics, Inc
FEI Number
1626313
Product Code
LLZ
Status
Terminated
Root Cause
Other
Initiated
February 9, 2007
Posted
April 29, 2011
Terminated
May 2, 2011
Address
3701 Kirby Dr, Ste 930, Houston, TX, 77098-3922

Description

Doctors Review System, OB-Windows/OB-View software release numbers 4.1 through 4.1.3, Digisonics, Inc., 3701 Kirby Drive, Houston, TX 77098 OB-Windows provides professional reporting and analysis of fetal growth throughout the gestational period, which can help evaluate clinical situations. OB-View provides image management for the user to capture, view, read, and edit images.

Reason

Study would on occasion have fetal heart rate and/or cervical length on the report that was not entered nor did it belong to that patient. Software displayed incorrect data for fetal heart rate and cervical length in OB ultrasound reports.

Action

The firm, Digisonics Inc, notified the affected customers of software release via e-mail on January 24, 2007 and follow-up e-mail on February 7, 2007. The e-mails described the product, problem and the actions taken. The firm was able to trap the error and after this report the developer was able to determine the cause and a new executable was sent and installed at the customers site. If you have any question, please call 800-940-3240 or e-mail: [email protected].

Distribution

Worldwide distribution: USA (nationwide) and country of Canada.

Quantity

128