Version: 1.2.0 (1.2.0+85ela410) Caption Health, Caption AI, Ultrasound Imaging System Application Software (GMDN: 40873), UDI: (01)00860003586303(10)1.2.0)

Recall: Z-2111-2021 · Initiated May 19, 2021

Recall

Recall Number
Z-2111-2021
Event Number
88101
Firm
Caption Health
FEI Number
3016665037
Product Code
QIH
Status
Terminated
Root Cause
Software design
Initiated
May 19, 2021
Terminated
January 31, 2023
Address
2000 Sierra Point Pkwy, Fl 8, Brisbane, CA, 94005-1845

Description

Version: 1.2.0 (1.2.0+85ela410) Caption Health, Caption AI, Ultrasound Imaging System Application Software (GMDN: 40873), UDI: (01)00860003586303(10)1.2.0)

Reason

The firm is aware of a issue with ultrasound software that results in unintended video frames being included. This could lead to an incorrect automated ejection fraction (AutoEF) and image quality score (IQS) estimates.

Action

On May 19, 2021 Caption health, Inc. contacted consignees via telephone to inform them about their recall. On May 20, 2021, the firm sent follow-up communication via via email. In addition to informing consignees about the recall, the firm also asked customers to take the following actions: 1. Until your system is updated, we recommend that all EF calculations are reviewed carefully using clinical expertise, to verify the accuracy of the results. 2. If you have any questions or concerns, please contact Caption health support line at 1-855-506-HELP between 7am-7pm ET. You can also email us at [email protected] at any time. We will follow-up with additional details relevant to your site by Monday May 24

Distribution

US : AZ, DE, NY, IL, CT OUS: None

Quantity

15 systems