FDA Recall Terminated

Brilliance CT Big Bore Oncology, model #728243; Intended to produce cross- sectional images of the body.

Recall: Z-2111-2016 · Initiated June 21, 2016

Recall

Recall Number
Z-2111-2016
Event Number
74511
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
June 21, 2016
Terminated
July 2, 2018
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

Brilliance CT Big Bore Oncology, model #728243; Intended to produce cross- sectional images of the body.

Reason

Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack and cause injury to patients, operators, by standers or service personnel.

Action

Philips is informing the consignee of this issue through: Customer Information Letter (CIL) 72800656 . International consignees will be contacted by Philips Healthcare Global Market representatives, who will distribute the CIL. A Field Service Engineer (FSE) will contact the customers to schedule a time to conduct the appropriate correction through Field Change Order (FCO) 72800656 once it is available. This FCO will address the above identified issue. A response form sent to the customer via certified mail. This package will also contain a return receipt.

Distribution

Distributed in China, Denmark, Germany, Hungary, Italy, Philippines, Spain, Sweden, Switzerland, Thailand, and United Kingdom.

Quantity

3 units