FDA Recall Terminated

Temno Coaxial Chiba Fine Needle Aspiration; a coaxial introducer needle (20G x 15 cm) with a Chiba biopsy needle (22G x 20 cm); sterile, for single use only sets, 5 sets per case; CareFusion, McGaw Park, IL 60085 USA, Made in Dominican Republic; REF CCH2220 For use in soft tissue aspiration such as breast, kidney, liver, lung, thyroid, lymph nodes and other various soft tissue masses. It is not intended for use in the bone.

Recall: Z-2111-2010 · Initiated June 25, 2010

Recall

Recall Number
Z-2111-2010
Event Number
56085
Firm
Carefusion 2200 Inc 1500 S Waukegan Rd Mpwm Bldg. Waukegan IL 60085-6728
FEI Number
3008002452
Product Code
GDF
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
June 25, 2010
Posted
July 28, 2010
Terminated
January 24, 2011

Description

Temno Coaxial Chiba Fine Needle Aspiration; a coaxial introducer needle (20G x 15 cm) with a Chiba biopsy needle (22G x 20 cm); sterile, for single use only sets, 5 sets per case; CareFusion, McGaw Park, IL 60085 USA, Made in Dominican Republic; REF CCH2220 For use in soft tissue aspiration such as breast, kidney, liver, lung, thyroid, lymph nodes and other various soft tissue masses. It is not intended for use in the bone.

Reason

The introducer needle is only 10 cm in length instead of the required 15 cm.

Action

Cardinal Health, on behalf of CareFusion, telephoned the customers on June 15, 2010 and sent a follow-up "URGENT: PRODUCT RECALL" letter on the same date via UPS next day air. The letter describes product, problem, and action taken by customers. The customers were instructed to immediately stop using affected product, return the enclosed acknowledgment form via fax to 847-686-9101 (note: Cardinal Health is required to confirm receipt of this notification from the customers), notify any customers to whom they may have distributed product to, contact CareFusion Customer Service at 800-653-6827 to arrange for credit and return of the affected product, and send the affected product to CareFusion, 1240 Waukegan Rd, Waukegan, IL 60085, Attn: Scott Fitzer, after obtaining an RMA/RGA number from CareFusion. If you have any questions, please contact 800-292-9332.

Distribution

Nationwide distribution: California, Florida, Georgia, Illinois, Kentucky, Massachusetts, New Jersey, New York and Pennsylvania.

Quantity

65 units