FDA Recall Terminated

Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.

Recall: Z-2110-2021 · Initiated June 4, 2021

Recall

Recall Number
Z-2110-2021
Event Number
88073
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
FRN
Status
Terminated
Root Cause
Software design
Initiated
June 4, 2021
Posted
August 6, 2021
Terminated
September 25, 2024
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.

Reason

System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.

Action

Baxter initiated the recall by letter on 06/04/2021. The letter described the problem and the hazard involved, and requested the consignee Baxter Technical Assistance in advance of making changes to your network or server infrastructure supporting the pumps or the Baxter Gateway, as indicated in Baxter's Gateway Server Installation Handoff document which customers receive upon installation. The firm is developing a software patch for all Spectrum IQ infusion pumps.

Distribution

United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.

Quantity

130135 units