Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
Recall
- Recall Number
- Z-2110-2021
- Event Number
- 88073
- Firm
- Baxter Healthcare Corporation
- FEI Number
- 1417572
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- June 4, 2021
- Posted
- August 6, 2021
- Terminated
- September 25, 2024
- Address
- 1 Baxter Pkwy, Deerfield, IL, 60015-4625
Description
Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
Baxter initiated the recall by letter on 06/04/2021. The letter described the problem and the hazard involved, and requested the consignee Baxter Technical Assistance in advance of making changes to your network or server infrastructure supporting the pumps or the Baxter Gateway, as indicated in Baxter's Gateway Server Installation Handoff document which customers receive upon installation. The firm is developing a software patch for all Spectrum IQ infusion pumps.
United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.
130135 units