Brennen Skin Graft Mesher. Surgical instrument designed to expand skin grafts'
Recall: Z-2110-2017
·
Initiated April 24, 2017
Recall
- Recall Number
- Z-2110-2017
- Event Number
- 77121
- Firm
- Molnlycke Health Care, Inc
- FEI Number
- 1629080
- Product Code
- FZW
- Status
- Open, Classified
- Root Cause
- Process design
- Initiated
- April 24, 2017
- Posted
- May 3, 2017
- Address
- 5550 Peachtree Pkwy, Ste 500, Norcross, GA, 30092-2555
Description
Brennen Skin Graft Mesher. Surgical instrument designed to expand skin grafts'
Reason
Sterilization validation failure.
Action
Consignees were notified by Urgent Safety Notice letter via certified mail on/about 4/24/2017.
Distribution
Domestic: U.S. Nationwide; Foreign: Austria, Australia, Bangladesh, Canada, Chile, Germany, Denmark, Finland, France, Great Britain, Hong Kong, Hungary, Ireland, India, Italy, Jordan, Korea, Kuwait, Malaysia, Norway, New Zealand, Peru, Saudi Arabia, Sweden Switzerland, Taiwan, Turkey, South Africa, United Arab Emirates.
Quantity
956 units