FDA Recall Open, Classified

REF 218-1604-SP-10, Profile 0 4mm AutoDrive Screw Sterile Qty 10, Rx Only, Sterile

Recall: Z-2108-2023 · Initiated May 31, 2023

Recall

Recall Number
Z-2108-2023
Event Number
92479
Firm
Osteomed, LLC
FEI Number
2027754
Product Code
GWO
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
May 31, 2023
Posted
July 7, 2023
Address
3885 Arapaho Rd, Addison, TX, 75001-4314

Description

REF 218-1604-SP-10, Profile 0 4mm AutoDrive Screw Sterile Qty 10, Rx Only, Sterile

Reason

Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole.

Action

On June 6, 2023, OsteoMed/Acumed issued an "Urgent: Medical Device Recall" notification to affected consignees via FedEx Ground. OsteoMed/Acumed asked consignees to take the following actions: IMMEDIATELY stop using and/or distributing the product and complete the following actions: 1) Perform a physical count of your inventory. Identify any inventory of the lot number listed above that is still in its packaging. Locate and remove the affected, unopened, unused packages of screws, such that they are out of service (i.e., quarantined). 2) Record this data (product and quantity) on the survey found at the end of this letter (the Outstanding Field Action Survey). 3) Complete the remaining, applicable fields in the Outstanding Field Action Survey and either return it via email to [email protected] or a physical copy can be mailed to: OsteoMed/Acumed, Attn: Lauren Freytag- R23-002-TX, 3885 Arapaho Rd, Addison, TX 75001 4) If you have product to return, please contact [email protected] so that an appropriate return number and return instructions can be provided. 5) If you do not have product to return, you are still required to complete the form below. 6) For patients implanted with these affected devices, OsteoMed/Acumed advises physicians to monitor patients.

Distribution

Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, KS, KY, MD, MO, MT, NE, NY, TN, TX and the country of SG.