FDA Recall Terminated

Maquet Getinge Group Operating Table System Magnus Carbon-Fibre Table Top Version 1180.16 The carbon-fibre MAGNUS table top (1180-16XX) serves to support and position patients for surgical procedures (e.g. cardiovascular surgery) and interventional examination immediately before, during and after the surgical phase. The table top is radiotranslucent and enables intraoperative use of X-ray equipment.

Recall: Z-2108-2012 · Initiated April 13, 2012

Recall

Recall Number
Z-2108-2012
Event Number
62590
Firm
Maquet Medical Systems, Usa
FEI Number
3001451475
Product Code
FQO
Status
Terminated
Root Cause
Device Design
Initiated
April 13, 2012
Posted
July 30, 2012
Terminated
September 4, 2013
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

Maquet Getinge Group Operating Table System Magnus Carbon-Fibre Table Top Version 1180.16 The carbon-fibre MAGNUS table top (1180-16XX) serves to support and position patients for surgical procedures (e.g. cardiovascular surgery) and interventional examination immediately before, during and after the surgical phase. The table top is radiotranslucent and enables intraoperative use of X-ray equipment.

Reason

As a result of a small number of customer complaints, MAQUET conducted an investigation and identified a potential problem with specific MAGNUS Carbon-Fibre Table Tops manufactured between February 23, 2010 and February 28, 2012 and distributed between November 11, 2010 and April 6, 2012. There is a potential issue with the motor brakes for the longitudinal shift drive, and the tilt drive.

Action

MAQUET GETINGE Group sent An "Urgent Device Field Correction" notification, dated 13 April 2012 via Federal Express to all affected customers. An acknowledgement form is included with the customer letter which can be mailed or faxed back to Maquet Quality by the notified customers. If the completed Acknowledgement form is not received within 15 days, up to three follow ups will be made and documented via telephone call to each customer. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use and notify their MAQUET Service Representative immediately. The MAQUET Service Representative will contact customers to make arrangements to replace both te logitudinal shift and tilt drives of their table top(s). For questions customers should call 888-627-8383, press option 2, followed by option 1. For questions regarding this recall call 201-995-8968.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of United Arab Emirates, Austria, Australia, Belgium, Brazil, Canada, Switzerland, China, Germany, Denmark, Spain, France, Hong Kong, Ireland, Israel, India, Italy, Japan, South Korea, Netherlands, Poland, Russia, Saudi Arabia, Sweden, Singapore, and Thailand.

Quantity

164 Units located in 27 Countries