Brand Name: Wolf Medical Product Name: 35 mL (cc) Luer Lock STERILE LATEX FREE NON PYROGENIC Model/Catalog Number: SYR35L Product Description: NON-Sterile syringes without needles for single use Component: No
Recall
- Recall Number
- Z-2106-2024
- Event Number
- 94520
- Firm
- Jiangsu Shenli Medical Production Co., Ltd. No.
- FEI Number
- 3003902757
- Product Code
- FMF
- Status
- Open, Classified
- Root Cause
- No Marketing Application
- Initiated
- April 5, 2024
- Posted
- May 20, 2024
- Address
- 20 Changzheng Road, Zhengluzhen Wujin Changzhou China
Description
Brand Name: Wolf Medical Product Name: 35 mL (cc) Luer Lock STERILE LATEX FREE NON PYROGENIC Model/Catalog Number: SYR35L Product Description: NON-Sterile syringes without needles for single use Component: No
Piston syringes sizes and configurations are out of the range of devices cleared under the firm's 510(k).
On 05/14/2024, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter informing customers that Jiangsu Shenli Medical Production Co., Ltd. is recalling piston syringes that were distributed without premarket clearance. The Recalling Firm stated that they received an FDA Warning Letter that the piston syringes were not legally marketed devices. Customers are instructed to: Examine their inventory and stop/cease use of all syringe products subject to this action and report the inventory quantity to Jiangsu Shenli Medical Production Co., Ltd. Considering the complicated return process, and high cost of returning the products, the firm is instructing customers to dispose of these recalled syringes in the U.S, using their facility s approved/established procedures for disposal of medical products, or by contacting a local third party company to assist with the disposal of the recalled syringes. Please report the destruction method and cost with Shenli before final implementation. After product disposal, please provide Shenli with confirmation of destruction of products. For questions or assistance - contact the general manager at email [email protected] or telephone number 0086-519-88731071
U.S. Nationwide distribution in the states of CA, FL, GA, IL, TN, and VA. After customers place US orders, then arrange production, then deliver to the cargos to customer designated port or warehouse, mostly in Shanghai port. The shipping company will ship the cargos to different destination such as Chicago, Los Angeles, Long beach, Miami, Savannah, Nashville, and Norfolk.
483210