Trilogy Evo, Evo Clinical Korean (1137823) and Evo Caregiver Korean (1137853) Instructions for Use
Recall
- Recall Number
- Z-2106-2023
- Event Number
- 92515
- Firm
- Philips Respironics, Inc.
- FEI Number
- 2518422
- Product Code
- CBK
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- May 25, 2023
- Posted
- July 7, 2023
- Address
- 1001 Murry Ridge Ln, Murrysville, PA, 15668-8517
Description
Trilogy Evo, Evo Clinical Korean (1137823) and Evo Caregiver Korean (1137853) Instructions for Use
There are errors present in the Korean Instructions For Use where the word "invasive," was incorrectly translated to "noninvasive".
Customers were mailed a recall notification dated 05/23/2023 identifying the reason for recall as a mistranslation from "invasive" to "non-invasive" in the Korean IFUs. Customers are to refer to the correct instructions as listed in the recall notification when referring to the updated IFU pages, keep a copy of the recall notification with the IFU, communicate the recall notification to all who need to be aware in their organization, and complete and return the provided Field Safety Notice Response Form. The response form can be returned to the Philips representative indicated in your specific recall notification.
Foreign Distribution: Korea.
1,581 units