FDA Recall Open, Classified

VersaCross RF Wire (VXW). Indicated for creation of an atrial septal defect in the heart. Product ID VXW0002 Device approved and commercialized in Japan only.

Recall: Z-2104-2025 · Initiated June 10, 2025

Recall

Recall Number
Z-2104-2025
Event Number
97036
Firm
Boston Scientific Corporation
FEI Number
2124215
Product Code
DXF
Status
Open, Classified
Root Cause
Packaging change control
Initiated
June 10, 2025
Posted
July 10, 2025
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700

Description

VersaCross RF Wire (VXW). Indicated for creation of an atrial septal defect in the heart. Product ID VXW0002 Device approved and commercialized in Japan only.

Reason

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

Action

On June 10, 2025, the firm began notifying consignees of the field action via Urgent Medical Device Removal letters. Customers are instructed to not use affected product. They should remove affected units from their inventory and segregate the units in a secure place until they can be returned to Boston Scientific.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.

Quantity

14,724