FDA Recall Open, Classified

VersaCross Access Solution (VXSK), Product Description and ID: VXSK (180J RF WIRE 63CM/45DEG/D0) VXSK0101 VXSK (180J RF WIRE 63CM/45DEG/D1) VXSK0102 VXSK (180J RF WIRE 63CM/55DEG/D0) VXSK0103 VXSK (180P RF WIRE 63CM/45DEG/D0) VXSK0121 VXSK (180P RF WIRE 63CM/45DEG/D1) VXSK0122 VXSK (180P RF WIRE 63CM/55DEG/D0) VXSK0123 VXSK (180P RF WIRE 63CM/55DEG/D1) VXSK0124 VXSK (180P RF WIRE 63CM/90DEG/D0) VXSK0125 VXSK (230J RF WIRE 63CM/45DEG/D0) VXSK0111 VXSK (230J RF WIRE 63CM/45DEG/D1) VXSK0112 VXSK (230P RF WIRE 63CM/45DEG/D0) VXSK0131 VXSK (230P RF WIRE 63CM/45DEG/D1) VXSK0132 VXSK (230P RF WIRE 63CM/55DEG/D0) VXSK0133 VXSK (230P RF WIRE 63CM/55DEG/D1) VXSK0134 VXSK (230P RF WIRE 81CM/45DEG/D0) VXSK0137 VXSK-180P RF WIRE 63CM/45DEG/D0 SHEATH VXSK0021 VXSK-180P RF WIRE 63CM/45DEG/D1 SHEATH VXSK0022 VXSK-180P RF WIRE 63CM/55DEG/D1 SHEATH VXSK0024 VXSK-180P RF WIRE 63CM/90DEG/D0 SHEATH VXSK0025 VXSK-230P RF WIRE 63CM/45DEG/D0 SHEATH VXSK0031 VXSK-230P RF WIRE 63CM/45DEG/D1 SHEATH VXSK0032 VXSK-230P RF WIRE 63CM/55DEG/D1 SHEATH VXSK0034

Recall: Z-2103-2025 · Initiated June 10, 2025

Recall

Recall Number
Z-2103-2025
Event Number
97036
Firm
Boston Scientific Corporation
FEI Number
2124215
Product Code
DXF
Status
Open, Classified
Root Cause
Packaging change control
Initiated
June 10, 2025
Posted
July 10, 2025
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700

Description

VersaCross Access Solution (VXSK), Product Description and ID: VXSK (180J RF WIRE 63CM/45DEG/D0) VXSK0101 VXSK (180J RF WIRE 63CM/45DEG/D1) VXSK0102 VXSK (180J RF WIRE 63CM/55DEG/D0) VXSK0103 VXSK (180P RF WIRE 63CM/45DEG/D0) VXSK0121 VXSK (180P RF WIRE 63CM/45DEG/D1) VXSK0122 VXSK (180P RF WIRE 63CM/55DEG/D0) VXSK0123 VXSK (180P RF WIRE 63CM/55DEG/D1) VXSK0124 VXSK (180P RF WIRE 63CM/90DEG/D0) VXSK0125 VXSK (230J RF WIRE 63CM/45DEG/D0) VXSK0111 VXSK (230J RF WIRE 63CM/45DEG/D1) VXSK0112 VXSK (230P RF WIRE 63CM/45DEG/D0) VXSK0131 VXSK (230P RF WIRE 63CM/45DEG/D1) VXSK0132 VXSK (230P RF WIRE 63CM/55DEG/D0) VXSK0133 VXSK (230P RF WIRE 63CM/55DEG/D1) VXSK0134 VXSK (230P RF WIRE 81CM/45DEG/D0) VXSK0137 VXSK-180P RF WIRE 63CM/45DEG/D0 SHEATH VXSK0021 VXSK-180P RF WIRE 63CM/45DEG/D1 SHEATH VXSK0022 VXSK-180P RF WIRE 63CM/55DEG/D1 SHEATH VXSK0024 VXSK-180P RF WIRE 63CM/90DEG/D0 SHEATH VXSK0025 VXSK-230P RF WIRE 63CM/45DEG/D0 SHEATH VXSK0031 VXSK-230P RF WIRE 63CM/45DEG/D1 SHEATH VXSK0032 VXSK-230P RF WIRE 63CM/55DEG/D1 SHEATH VXSK0034

Reason

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

Action

On June 10, 2025, the firm began notifying consignees of the field action via Urgent Medical Device Removal letters. Customers are instructed to not use affected product. They should remove affected units from their inventory and segregate the units in a secure place until they can be returned to Boston Scientific.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.

Quantity

14,724