FDA Recall Terminated

Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment plan verification in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.

Recall: Z-2100-2017 · Initiated February 23, 2015

Recall

Recall Number
Z-2100-2017
Event Number
77177
Firm
Mobius Medical Systems, LP
FEI Number
3008610750
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
February 23, 2015
Posted
May 16, 2017
Terminated
July 10, 2017
Address
4615 Southwest Fwy, Ste 330, Houston, TX, 77027-7106

Description

Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment plan verification in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.

Reason

Mobius3D version 1.5.0 contained a defect in software code which affects users who perform beam customization, and may lead to a discrepancy in dose calculation between this version of Mobius3D (1.5.0) and the prior version (1.4.2), where none should have been expected.

Action

The recalling firm sent an Advisory Notice letter dated February 23, 2015 to customers. The letter identified the affected product, problem and actions to be taken. For questions contact Mobius Medical Systems Customer Support at [email protected].

Distribution

Worldwide distribution - US Nationwide and the countries of Sweden, Canada, Netherlands, Denmark, Spain, Switzerland, Australia, Austria, Israel, United Kingdom, France

Quantity

270