Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment plan verification in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
Recall
- Recall Number
- Z-2100-2017
- Event Number
- 77177
- Firm
- Mobius Medical Systems, LP
- FEI Number
- 3008610750
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 23, 2015
- Posted
- May 16, 2017
- Terminated
- July 10, 2017
- Address
- 4615 Southwest Fwy, Ste 330, Houston, TX, 77027-7106
Description
Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment plan verification in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
Mobius3D version 1.5.0 contained a defect in software code which affects users who perform beam customization, and may lead to a discrepancy in dose calculation between this version of Mobius3D (1.5.0) and the prior version (1.4.2), where none should have been expected.
The recalling firm sent an Advisory Notice letter dated February 23, 2015 to customers. The letter identified the affected product, problem and actions to be taken. For questions contact Mobius Medical Systems Customer Support at [email protected].
Worldwide distribution - US Nationwide and the countries of Sweden, Canada, Netherlands, Denmark, Spain, Switzerland, Australia, Austria, Israel, United Kingdom, France