19" Barco MFCD 1219 (Touch Screen) + Low Profile Stand and Rack with LCD Arm Agfa's Computed Radiography Systems with NX2.X software are indicated for use in providing diagnostic quality images to aid the physician with diagnosis.
Recall
- Recall Number
- Z-2100-2010
- Event Number
- 56158
- Firm
- AGFA Corp.
- FEI Number
- 3001236302
- Product Code
- MQB
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 30, 2010
- Posted
- July 26, 2010
- Terminated
- April 28, 2011
- Address
- 10 S Academy St, Greenville, SC, 29601-2632
Description
19" Barco MFCD 1219 (Touch Screen) + Low Profile Stand and Rack with LCD Arm Agfa's Computed Radiography Systems with NX2.X software are indicated for use in providing diagnostic quality images to aid the physician with diagnosis.
Monitor fell from the mounting bracket used to support the monitor.
The firm, AGFA Healthcare sent an 'URGENT SAFETY NOTICE" dated June 30, 2010, via FED-EX to customers. The letter describes product, problem and actions to be taken by the firm and customers. AGFA will provide a safety check to the customers facility, have a service representative visit , in the near future, their site to inspect the monitor/s and attach a warning label on each monitor checked for preventative action. The customers are ask to distribute the information within their facility to all individuals who need to be aware, to complete the attached URGENT SAFETY NOTICE FEEDBACK FORM as soon as possible and return via fax at 864-421-1664. If you have any questions about this matter, please feel free to contact me via email at [email protected] or call me at 864-421-1754.
Worldwide distribution: USA and countries of Puerto Rico and Canada.
805 units