FDA Recall Terminated

Safety PICC Insertion Tray with Chloraprep OneStep - Model #5026CL-9 - 12 pr case. Packaged for Physician use only.

Recall: Z-2096-2015 · Initiated November 5, 2014

Recall

Recall Number
Z-2096-2015
Event Number
69762
Firm
Custom Healthcare Systems, Inc.
FEI Number
1121091
Product Code
GCB
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
November 5, 2014
Posted
July 17, 2015
Terminated
January 12, 2016
Address
4205 Eubank Rd, Richmond, VA, 23231-4328

Description

Safety PICC Insertion Tray with Chloraprep OneStep - Model #5026CL-9 - 12 pr case. Packaged for Physician use only.

Reason

An INCORRECT item number is packed within the OUTSIDE "PIGGYBACKED" PACKAGE. The interior sealed STERILE KIT is correct and NO adverse issues affect that part of the item. The NON-STERILE OUTSIDE "PIGGYBACKED" PACKAGE mistakenly contains ITEM #8881579121 - Syringe prefilled, Saline Flush, 09% Sodium Chloride and is a NONSTERILE PRODUCT that CANNOT BE DROPPED INTO THE STERILE FIELD.

Action

Custom Healthcare System sent an Urgent Product Recall letter dated November 5, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. PLEASE CHECK YOUR INVENTORY AND QUARANTINE ANY AFFECTED PRODUCT MATCHING THE DESCRIPTIONS ABOVE. Call 804 (421-5959) with regards to returning the affected products. Return Authorizations will be issued and product pickup and return will be scheduled.

Distribution

US Distribution to CT.

Quantity

3 Cs-Lot#92414 , 25 Cs-Lot#100214, 24 Cs-Lot#100914