TrueDent Yellow, TDM140, Model No. OBJ-09169 Light-curable methacrylate-based resin that enables fabrication of dental appliances
Recall
- Recall Number
- Z-2093-2025
- Event Number
- 97021
- Firm
- STRATASYS LTD
- FEI Number
- 3017374693
- Product Code
- EBI
- Status
- Completed
- Root Cause
- Software design
- Initiated
- January 19, 2024
- Posted
- July 7, 2025
- Address
- 1 Holtzman St, Rehovot Israel
Description
TrueDent Yellow, TDM140, Model No. OBJ-09169 Light-curable methacrylate-based resin that enables fabrication of dental appliances
Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
The firm initiated the recall to the consignee level on January 15, 2024. Customers were asked to coordinate a cartridge replacement. The reporting of this recall was done by the firm retroactively; the issue was resolved in January 2024.
US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, Nebraska, New York, North Carolina, North Dakota, Oklahoma, Oregon, Utah, Wisconsin.
68