FDA Recall Terminated

ST-AIA PACK Prog II; Part Number: 025239 Assay, Reproductive Hormone

Recall: Z-2091-2018 · Initiated March 5, 2018

Recall

Recall Number
Z-2091-2018
Event Number
79813
Firm
Tosoh Bioscience Inc
FEI Number
3005529799
Product Code
JLS
Status
Terminated
Root Cause
Device Design
Initiated
March 5, 2018
Terminated
August 31, 2020
Address
3600 Gantz Rd, Grove City, OH, 43123-1895

Description

ST-AIA PACK Prog II; Part Number: 025239 Assay, Reproductive Hormone

Reason

Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.

Action

The firm TOSOH Bioscience Inc., sent an "URGENT: MEDICAL DEVICE RECALL" letter dated March 23, 2018 to affected customers. The letter described the product, problem and action to be taken. The customers were instructed to do the following: *Based on our revised labeling, the assays identified in Table 1 should not be used for patients who are being treated with Asfotase Alfa. " For patients that are being treated with Asfotase Alfa, use an alternative test method that does not utilize alkaline phosphatase technology. " Inform all medical professionals that assays utilizing alkaline phosphatase-based technologies must not be used for patients receiving Asfotase Alfa treatment. Continue to use the assays listed in Table 1 for any patients who are not receiving Asfotase Alfa treatment. " Complete and return the attached Acknowledgement Form to Tosoh Bioscience, Inc., within 15-days of receiving this notification via Fax: 1-650-636-8651, Email: [email protected] or Regular mail: 3600 Gantz Rd., Grove City, OH 43123. " Maintain this notification with your laboratory records and forward this information to others who may have received this product. " If you have obtained unexpected test results or received any complaints of illness or adverse events associated with the use of the products identified in the table above, contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. Should you have any questions regarding this medical device recall, please feel free to call (800) 248-6764 or email: [email protected]. Monday - Friday, from 9:00 AM to 5:00 PM (PST).

Distribution

Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.

Quantity

5990