Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number:DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number:DYNJ56436B; 3) SCC RF PAIN MGMT, Model Number:DYNJ56436C; 4) CPNB SETUP TRAY, Model Number:DYNJRA1979; 5) CPNB SETUP TRAY, Model Number:DYNJRA1979A; 6) CPNB SETUP TRAY, Model Number:DYNJRA1979B; 7) VARITHENA TRAY 2, Model Number:DYNJRA2608; 8) CPNB SETUP TRAY, Model Number:SAMPA0108
Recall
- Recall Number
- Z-2090-2026
- Event Number
- 98642
- FEI Number
- 1417592
- Product Code
- OFT
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- January 7, 2026
- Posted
- May 7, 2026
- Address
- 3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753
Description
Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number:DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number:DYNJ56436B; 3) SCC RF PAIN MGMT, Model Number:DYNJ56436C; 4) CPNB SETUP TRAY, Model Number:DYNJRA1979; 5) CPNB SETUP TRAY, Model Number:DYNJRA1979A; 6) CPNB SETUP TRAY, Model Number:DYNJRA1979B; 7) VARITHENA TRAY 2, Model Number:DYNJRA2608; 8) CPNB SETUP TRAY, Model Number:SAMPA0108
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/14/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-
Worldwide - US Nationwide distribution.
1774 units