FDA Recall Terminated

Arietta 70 Ultrasound System Intended for use by trained personnel (doctor, sonographer. etc.) for the diagnostic Ultrasound evaluation.

Recall: Z-2090-2017 · Initiated May 2, 2017

Recall

Recall Number
Z-2090-2017
Event Number
77051
Firm
Hitachi Medical Systems America Inc
FEI Number
1528028
Product Code
IYN
Status
Terminated
Root Cause
Device Design
Initiated
May 2, 2017
Terminated
November 19, 2018
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087-2371

Description

Arietta 70 Ultrasound System Intended for use by trained personnel (doctor, sonographer. etc.) for the diagnostic Ultrasound evaluation.

Reason

The strength of the arm mounting mechanism was insufficient and had to be redesigned.

Action

An Urgent: Device Correction Notice was sent via overnight carrier on 5/5/17. The Notice identified the affected product and problem. The firm will replace the Monitor Arm to correct the issue at no cost to the customer. If the problem occurs, customers are to call 800-800-4925 for immediate service. Hitachi Service will update the hardware at each site. Customers can contact [email protected] with questions, comments, or requests regarding the Notice.

Distribution

Distributed to the states of: AZ, CA, FL, HI, IL MA, MD, ME, MN, MO, NC, NJ, NY, OH, PA, SC, TN, and TX.

Quantity

77