Arietta 70 Ultrasound System Intended for use by trained personnel (doctor, sonographer. etc.) for the diagnostic Ultrasound evaluation.
Recall
- Recall Number
- Z-2090-2017
- Event Number
- 77051
- Firm
- Hitachi Medical Systems America Inc
- FEI Number
- 1528028
- Product Code
- IYN
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- May 2, 2017
- Terminated
- November 19, 2018
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087-2371
Description
Arietta 70 Ultrasound System Intended for use by trained personnel (doctor, sonographer. etc.) for the diagnostic Ultrasound evaluation.
The strength of the arm mounting mechanism was insufficient and had to be redesigned.
An Urgent: Device Correction Notice was sent via overnight carrier on 5/5/17. The Notice identified the affected product and problem. The firm will replace the Monitor Arm to correct the issue at no cost to the customer. If the problem occurs, customers are to call 800-800-4925 for immediate service. Hitachi Service will update the hardware at each site. Customers can contact [email protected] with questions, comments, or requests regarding the Notice.
Distributed to the states of: AZ, CA, FL, HI, IL MA, MD, ME, MN, MO, NC, NJ, NY, OH, PA, SC, TN, and TX.
77