ADVIA 120/2120/2120i 3-in-1 TESTpoint Control (ABN2 Control)
Recall
- Recall Number
- Z-2089-2025
- Event Number
- 97037
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2432235
- Product Code
- JCN
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- June 4, 2025
- Posted
- July 3, 2025
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005
Description
ADVIA 120/2120/2120i 3-in-1 TESTpoint Control (ABN2 Control)
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
An Urgent Medical Device Correction notification dated 6/4/25 was mailed to consignees. This notification provides consignees with calibration instructions given the issue found in affected lots. Additionally, instructions are given on how to proceed with product depending on performance. Consignees are to discontinue use of and discard of the calibrator SETPOINT lots impacted by this recall. Consignee locations are to review the provided notification with their Medical Directors to determine the appropriate course of action. A copy of the recall notification should be retained in laboratory records and shared with users. Consignees are to complete and return the provided Effectiveness Check Form within 30 days of receipt of the recall notification. Consignees with any questions can contact their Siemens Customer Care Center or local Siemens Healthineers technical support representative.
Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.
3,266 units