FDA Recall Terminated

Cook Medical Ciaglia Blue Dolphin Percutaneous Tracheostomy Introducer Tray, G51971, C-PTBSYJC-2600-PERC6.

Recall: Z-2089-2010 · Initiated April 21, 2010

Recall

Recall Number
Z-2089-2010
Event Number
56084
Firm
Cook, Inc.
FEI Number
1820334
Product Code
JOH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 21, 2010
Posted
July 26, 2010
Terminated
January 23, 2012
Address
750 Daniels Way, Bloomington, IN, 47404-9120

Description

Cook Medical Ciaglia Blue Dolphin Percutaneous Tracheostomy Introducer Tray, G51971, C-PTBSYJC-2600-PERC6.

Reason

These set and/or tray products are being recalled because they include a Covidien Shiley 6PERC or Covidien Shiley 8PERC tracheostomy tube component that has been associated with cuff leakage as a result of the pilot balloon inflation assembly.

Action

Cook Medical sent an Urgent Product Recall letter dated April 21, 2010, to hospitals and other user facilities receiving product direct from Cook as well as distribution centers receiving direct shipments from Cook. Consignees were directed to cease distribution of product and notify their customers immediately.

Distribution

Worldwide distribution: USA, Canada, Mexico, Australia, Chile, Costa Rica, Denmark, Dominican Republic, Guatemala, India, Malaysia, Panama, Peru, Puerto Rico, Republic of Korea, Trinidad and Tobago, and Uruguay.

Quantity

28819 total, all products