FDA Recall Open, Classified

ADVIA 120/2120/2120i 3-in-1 TESTpoint Control (Norm Control)

Recall: Z-2088-2025 · Initiated June 4, 2025

Recall

Recall Number
Z-2088-2025
Event Number
97037
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
JCN
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
June 4, 2025
Posted
July 3, 2025
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

ADVIA 120/2120/2120i 3-in-1 TESTpoint Control (Norm Control)

Reason

Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.

Action

An Urgent Medical Device Correction notification dated 6/4/25 was mailed to consignees. This notification provides consignees with calibration instructions given the issue found in affected lots. Additionally, instructions are given on how to proceed with product depending on performance. Consignees are to discontinue use of and discard of the calibrator SETPOINT lots impacted by this recall. Consignee locations are to review the provided notification with their Medical Directors to determine the appropriate course of action. A copy of the recall notification should be retained in laboratory records and shared with users. Consignees are to complete and return the provided Effectiveness Check Form within 30 days of receipt of the recall notification. Consignees with any questions can contact their Siemens Customer Care Center or local Siemens Healthineers technical support representative.

Distribution

Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.

Quantity

3,435 units