RadiForce GX540, 5MP Monochrome LCD Monitor, Part Code No. 0FTD1826A5, 0FTD1826B5, 0FTD1826C5, UPC 690592034752. This product is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.
Recall
- Recall Number
- Z-2087-2015
- Event Number
- 71628
- Firm
- Eizo Corporation 153 Shimokashiwano Hakusanshi Japan
- FEI Number
- 3000237660
- Product Code
- PGY
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 7, 2015
- Posted
- July 16, 2015
- Terminated
- November 30, 2016
Description
RadiForce GX540, 5MP Monochrome LCD Monitor, Part Code No. 0FTD1826A5, 0FTD1826B5, 0FTD1826C5, UPC 690592034752. This product is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.
EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular graphics boards with DisplayPort.
EIZO sent An Urgent Field Safety Notice dated July 7, 2015, to all customers who purchased the RadiForce RX440, RX650, RX850 and GX540. The notice informs customers that the RadiForce monitors could display abornormally in 6-bit (64 gradations) instead of normal 8-bit (256 gradations) or 10-bit (1024 gradations) while connecting to particular graphics boards with DisplayPort. The notice provides the customers with the actions to be taken and contact information (1-800-800-5202, M-F, 8am-4pm, PST, [email protected], 5710 Warland Drive, Cypress, CA 90630).
Nationwide Distribution including OH, FL, IL, TX, OH, NY, AZ, and NJ.
19 units