FDA Recall Terminated

RadiForce GX540, 5MP Monochrome LCD Monitor, Part Code No. 0FTD1826A5, 0FTD1826B5, 0FTD1826C5, UPC 690592034752. This product is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.

Recall: Z-2087-2015 · Initiated July 7, 2015

Recall

Recall Number
Z-2087-2015
Event Number
71628
Firm
Eizo Corporation 153 Shimokashiwano Hakusanshi Japan
FEI Number
3000237660
Product Code
PGY
Status
Terminated
Root Cause
Device Design
Initiated
July 7, 2015
Posted
July 16, 2015
Terminated
November 30, 2016

Description

RadiForce GX540, 5MP Monochrome LCD Monitor, Part Code No. 0FTD1826A5, 0FTD1826B5, 0FTD1826C5, UPC 690592034752. This product is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.

Reason

EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular graphics boards with DisplayPort.

Action

EIZO sent An Urgent Field Safety Notice dated July 7, 2015, to all customers who purchased the RadiForce RX440, RX650, RX850 and GX540. The notice informs customers that the RadiForce monitors could display abornormally in 6-bit (64 gradations) instead of normal 8-bit (256 gradations) or 10-bit (1024 gradations) while connecting to particular graphics boards with DisplayPort. The notice provides the customers with the actions to be taken and contact information (1-800-800-5202, M-F, 8am-4pm, PST, [email protected], 5710 Warland Drive, Cypress, CA 90630).

Distribution

Nationwide Distribution including OH, FL, IL, TX, OH, NY, AZ, and NJ.

Quantity

19 units