Continuous Ventilator/Continuous Ventilator with Respiratory Mechanics, Model No. V1000
Recall
- Recall Number
- Z-2084-2013
- Event Number
- 48997
- Firm
- Respironics California Inc
- FEI Number
- 2518422
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- October 29, 2007
- Posted
- August 26, 2013
- Terminated
- August 26, 2013
- Address
- 2271 Cosmos Ct, Carlsbad, CA, 92011-1517
Description
Continuous Ventilator/Continuous Ventilator with Respiratory Mechanics, Model No. V1000
The power to the Esprit Display's Backlight is interrupted, causing the GUI to flicker or go dark and making the display unreadable.
A Medical Device Field Correction letter dated 10/26/2007 was sent to all customers who received the Esprit Ventilator. The recall letter informed the customers of the problems identified and the actions to be taken. Customers with questions are instructed to contact Respironics' Customer Service Dept. at (800) 345-6443, select Option#5, then select Option #4.
USA Nationwide and Worldwide to Algeria, Argentina, Australia, Bangladesh, Brazil, Cambodia, Canada, Chile, Colombia, Ecuador, Egypt, France, Germany, Guatemala, Hong Kong, India, Italy, Japan, Libya, Malaysia, Mexico, Pakistan, Peru, Philippines, Puerto Rico, Russia, South Africa, South Korea, Spain, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Vietnam.
3,255 units (2302 units in US)