FDA Recall Open, Classified

Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

Recall: Z-2083-2025 · Initiated May 14, 2025

Recall

Recall Number
Z-2083-2025
Event Number
96943
Firm
Edwards Lifesciences, LLC
FEI Number
2015691
Product Code
DQR
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 14, 2025
Posted
July 10, 2025
Address
1 Edwards Way, Irvine, CA, 92614-5688

Description

Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

Reason

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

Action

On 05/16/2025, and 05/19/2025, the firm sent via FedEx overnight an "URGENT: PRODUCT RECALL" Letter to customers informing them that Edwards has confirmed a small number of occurrences impacting the OptiSite Arterial Perfusion cannula in which a 3mm to 4mm section of wire, from the wire-reinforcement coil at the cannula tip, was found to be released from the cannula body. The risk for the wire-reinforcement coil protruding out of the cannula body is major tissue damage and hemolysis. Customers are instructed to: " Verify their inventory on the attached customer acknowledgement form. " Share this notice with appropriate clinical staff at their site. " No patient follow-up or notification is necessary. " Return a completed Customer Acknowledgment Form to Edwards eCV Representative or via email to Edwards Customer Service at [email protected] within 15 days of receipt of this notification. " After receiving RGA number, return any impacted product to Edwards at the address below. A credit will be issued upon receipt of returned product. enableCV Distribution Center 6644 W. 2100 S. Suite D West Valley, UT 84128 For questions/assistance contact - Customer Service at 1-888-943-2783, or email: [email protected]

Distribution

Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, Puerto Rico, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN and the countries of Albania, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Netherlands, Northern Ireland, Norway, Pakistan, Panama, Poland, Portugal, Romania, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates (UAE), United Kingdom.

Quantity

847