FDA Recall Terminated

RECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during proximal preparation.

Recall: Z-2083-2014 · Initiated June 3, 2014

Recall

Recall Number
Z-2083-2014
Event Number
68440
Firm
DePuy Orthopaedics, Inc.
FEI Number
1818910
Product Code
KWA
Status
Terminated
Root Cause
Device Design
Initiated
June 3, 2014
Posted
July 21, 2014
Terminated
May 17, 2017
Address
700 Orthopaedic Dr, Warsaw, IN, 46582-3994

Description

RECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during proximal preparation.

Reason

The product can be difficult to remove from the Distal Stem both out of the package and after proximal reaming.

Action

DePuy Synthes sent an Urgent Information - Device Correction Notice letter dated June 24, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: 1. Immediate Remedy: Discard the RECLAIM Taper Sleeve Protector instrument before stem implantation. 2. Changes to the literature and IFU: Documents are revised as follows: " RECLAIM Revision Hip Systems Surgical Technique # DSUS/JRC/0614/0200 pages 11, 12, 13, 16, 30, 34, 35 (figure numbering from pages 11 to 27 were also revised): Removed information regarding taper sleeve protector. " RECLAIM Revision Hip Systems Surgical Technique FAQ # DSUS/JRC/0614/0202 pages 1 and 2: Removed information regarding taper sleeve protector. " RECLAIM Revision Hip Systems IFU #090200799 REV E Page 5: Verbiage will be revised and implemented as soon as possible, depending on regulatory approvals. 3. Change: The company consulted with design surgeons and developed a strategy to eliminate the taper sleeve protector from the system. As a result of removing the taper sleeve protector, the company is evaluating whether there is a need for a protective component to be used during the distribution Customers with questions were instructed to call 574-371-4917. For questions regarding this recall call 574-371-4577.

Distribution

Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI & WV and Internationally to AUSTRALIA, AUSTRIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, IBERIA, IRELAND, ISRAEL, ITALY, NEW ZEALAND, NORWAY, POLAND, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWITZERLAND & UK.

Quantity

10,467 units