FDA Recall Terminated

IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking System in a radiotherapy environment.

Recall: Z-2082-2019 · Initiated July 18, 2019

Recall

Recall Number
Z-2082-2019
Event Number
83358
Firm
Medical Intelligence Medizintechnik Gmbh Robert-bosch-str. 8 Schwabmunchen Germany
FEI Number
3003251135
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
July 18, 2019
Terminated
August 11, 2023

Description

IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking System in a radiotherapy environment.

Reason

Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.

Action

The firm sent an Urgent Important Field Safety Notification letter dated July 2019 via email. The identified the affected product, problem and actions to be taken. Customers were instructed to post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed and advise the appropriate personnel, working with this product, on the content of this letter.

Distribution

Worldwide Distribution - US Nationwide

Quantity

496 units