FDA Recall Terminated

HDL Cholesterol Reagent (HDL 2 x 200), Catalog No. A15625. Intended for quantitative determination of HDL cholesterol in the high density lipoprotein fraction of human serum or plasma.

Recall: Z-2082-2017 · Initiated April 3, 2017

Recall

Recall Number
Z-2082-2017
Event Number
76962
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
LBS
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
April 3, 2017
Terminated
May 31, 2018
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

HDL Cholesterol Reagent (HDL 2 x 200), Catalog No. A15625. Intended for quantitative determination of HDL cholesterol in the high density lipoprotein fraction of human serum or plasma.

Reason

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.

Action

An Urgent Medical Device Recall letter dated 4/3/17 was sent to customers to inform them that Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole, Sulfasalazine, and Sulfapyridine. Customers are informed that with the administration of these drugs, the effects to patient analytes are: - Metamizole (Dipyrone) may generate false low Cholesterol, Enzymatic Creatinine, HDL Cholesterol, Lactate, Triglycerides and Uric Acid results. -Sulfasalazine may generate false high Salicylate results. -Sulfapyridine may generate false low Salicylate results. Customers are informed of the action and resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding the notice are instructed to contact Customer Support Center -Via http://www.beckmancoulter.com/customersupport/support -Via phone, call 1-800-854-3633 in the United States and Canada Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.

Distribution

Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States including Puerto Rico, and Vietnam.

Quantity

91,996 units total (51,027 units in US)