FDA Recall
Terminated
CENTURION Vision System
Recall: Z-2081-2016
·
Initiated May 24, 2016
Recall
- Recall Number
- Z-2081-2016
- Event Number
- 74282
- Firm
- Alcon Research, Ltd.
- FEI Number
- 1610287
- Product Code
- HQC
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 24, 2016
- Posted
- June 28, 2016
- Terminated
- April 13, 2017
- Address
- 6201 South Fwy Fort, Worth, TX, 76134-2099
Description
CENTURION Vision System
Reason
The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed.
Action
Alcon has already notified the affected customer verbally on May 24, 2016, and via written correspondence on June 1, 2016. The device was corrected on May 26, 2016, and the Power Entry Switch that was removed from the customers CENTURION Vision System did not exhibit a grounding pin susceptible to detachment. For further questions, please call (817) 293-0450.
Distribution
US Distribution to Illinois
Quantity
1 unit