Lactate Reagent (LACT 2 x 50), Catalog No. A95550 Intended for the quantitative determination of Lactate concentration in human plasma and cerebrospinal fluid (CSF).
Recall
- Recall Number
- Z-2078-2017
- Event Number
- 76962
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- CFJ
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- April 3, 2017
- Terminated
- May 31, 2018
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
Lactate Reagent (LACT 2 x 50), Catalog No. A95550 Intended for the quantitative determination of Lactate concentration in human plasma and cerebrospinal fluid (CSF).
Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
An Urgent Medical Device Recall letter dated 4/3/17 was sent to customers to inform them that Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole, Sulfasalazine, and Sulfapyridine. Customers are informed that with the administration of these drugs, the effects to patient analytes are: - Metamizole (Dipyrone) may generate false low Cholesterol, Enzymatic Creatinine, HDL Cholesterol, Lactate, Triglycerides and Uric Acid results. -Sulfasalazine may generate false high Salicylate results. -Sulfapyridine may generate false low Salicylate results. Customers are informed of the action and resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding the notice are instructed to contact Customer Support Center -Via http://www.beckmancoulter.com/customersupport/support -Via phone, call 1-800-854-3633 in the United States and Canada Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.
Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States including Puerto Rico, and Vietnam.
92,837 units total (71,271 units in US)