Siemens syngo.plaza. Radiological image processing system
Recall
- Recall Number
- Z-2078-2012
- Event Number
- 62462
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 24, 2012
- Posted
- July 26, 2012
- Terminated
- January 14, 2014
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355
Description
Siemens syngo.plaza. Radiological image processing system
Siemens became aware of a potential malfunction when using syngo.plaza with software version VA20C_HF01. For datasets with distance measurements it can happen, that all images are not loaded to Viewer but no error message is displayed to user.
Siemens sent a Customer Safety Advisory Notice letter dated February 24, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to ensure that the prefix defined in the viewer settings for "Content Creator's Initials" is not empty. Then all images will be display correctly. This has to be checked for every user. Further questions please call 610-219-6300.
(USA) nationwide including the states of CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.
2