FDA Recall Terminated

Siemens syngo.plaza. Radiological image processing system

Recall: Z-2078-2012 · Initiated February 24, 2012

Recall

Recall Number
Z-2078-2012
Event Number
62462
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
February 24, 2012
Posted
July 26, 2012
Terminated
January 14, 2014
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

Siemens syngo.plaza. Radiological image processing system

Reason

Siemens became aware of a potential malfunction when using syngo.plaza with software version VA20C_HF01. For datasets with distance measurements it can happen, that all images are not loaded to Viewer but no error message is displayed to user.

Action

Siemens sent a Customer Safety Advisory Notice letter dated February 24, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to ensure that the prefix defined in the viewer settings for "Content Creator's Initials" is not empty. Then all images will be display correctly. This has to be checked for every user. Further questions please call 610-219-6300.

Distribution

(USA) nationwide including the states of CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.

Quantity

2