FDA Recall Terminated

U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract dysfunctions

Recall: Z-2076-2019 · Initiated July 17, 2019

Recall

Recall Number
Z-2076-2019
Event Number
83310
Firm
LABORIE MEDICAL TECHNOLOGIES, CORP
FEI Number
3017448360
Product Code
FEN
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 17, 2019
Terminated
July 10, 2023
Address
400 Avenue D, Ste 10, Williston, VT, 05495-7828

Description

U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract dysfunctions

Reason

Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard

Action

Laborie issued letter via FedX on July 17, 2019 advising of the problem, health risk, and action to take: Quarantine the product and unplug the power supply from the wall socket until a replacement is provided. A new power supply unit(s) will be provided along with instructions on how to replace the power supplie in the field. Contact Laborie/MMS customer care with any questions, concerns or problems: Laborie MMS Customer Care Address: 400 Avenue D, Suite 10, Williston, VT 05495 Phone Number: 800.522.6743 Email: [email protected]

Distribution

US distribution in AZ, and PA.

Quantity

2 US