Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autostainer, Model S3400; and Autostainer Plus, Model S3800.

Recall: Z-2074-2020 · Initiated November 5, 2019

Recall

Recall Number
Z-2074-2020
Event Number
84465
Firm
Dako Denmark A/S Produktionsvej 42 Glostrup Denmark
FEI Number
3002806778
Product Code
KPA
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 5, 2019
Posted
May 21, 2020
Terminated
May 17, 2022

Description

Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autostainer, Model S3400; and Autostainer Plus, Model S3800.

Reason

The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining.

Action

The firm issued letters dated 10/30/2019 via email or standard mail, or phone call beginning 11/5/2019 notifying the customer of the issue, providing the risk to health, and actions that needed to be taken. On 09/03/2020, Agilent Dako sent Urgent Field Safety Notices to customers who were asked to take the following actions: 1) Carefully read and communicate to those within your organization of the updates stated in this notification letter and attached Autostainer Link Basic User Guide (rev. 06). Your local sales team will confirm with you that you have read and acknowledged this notification. 2) Your Agilent representative should be contacted immediately for assistance in the event of leakage. 3) Please discard previous versions of the Autostainer Link Basic User Guide that you are currently using.

Distribution

Distribution was nationwide. There was government/military distribution. Foreign distribution was made to Canada, Australia, Bangaladesh, Belarus, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Macedonia, Morocco, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Poland, Qatar, Romania, Russia, Saudi Arabia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Uruguay, and Vietnam.

Quantity

796 devices