Access BR Monitor, Part No. 387620
Recall
- Recall Number
- Z-2074-2017
- Event Number
- 77029
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- MOI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 13, 2017
- Terminated
- June 4, 2018
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
Access BR Monitor, Part No. 387620
Following biases observed in the results of the French National Quality Controls, Beckman Coulter has been informed that a study completed by the French Competent Authority Agence nationale de scurit du Medicament et des produits de sant (ANSM) showed diagnosed cancer patients in non-remission of their metastatic disease, may have results below the upper reference limit (URL) of 31.3 U/mL as stated in the Access BR Monitor instructions for use (IFU).
An urgent Medical Device Recall letter dated 4/13/17 was sent to customers following biases observed in the results of the French National Quality Controls, Beckman Coulter has been informed that a study completed by the French Competent Authority Agence nationale de scurit du Medicament et des produits de sant (ANSM) showed diagnosed cancer patients in non-remission of their metastatic disease, may have results below the upper reference limit (URL) of 31.3 U/mL as stated in the Access BR Monitor instructions for use (IFU). Customers are informed of the impact and the actions to be taken. Customers with any questions regarding the notice are instructed to contact Customer Technical Support Center: From our website: http://www.beckmancoulter.com By phone: call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter representative.
Affected geographies: Algeria, Australia, Austria, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Libya, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Sierra Leone, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Venezuela, Vietnam, Zimbabwe
45,278 units total (2,419 units in US)