FDA Recall Terminated

CapsoCam Plus *** System, Imaging, Gastrointestinal, Wireless, Capsule, Electrical Rating 3.1V DC/10mA REF 2795 - Product Usage: The CapsoCam Plus video capsule system is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.

Recall: Z-2073-2020 · Initiated July 3, 2019

Recall

Recall Number
Z-2073-2020
Event Number
85532
Firm
Capso Vision, Inc.
FEI Number
3008062894
Product Code
NEZ
Status
Terminated
Root Cause
Process control
Initiated
July 3, 2019
Terminated
January 28, 2021
Address
18805 Cox Ave, Ste 250, Saratoga, CA, 95070-6615

Description

CapsoCam Plus *** System, Imaging, Gastrointestinal, Wireless, Capsule, Electrical Rating 3.1V DC/10mA REF 2795 - Product Usage: The CapsoCam Plus video capsule system is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.

Reason

Possible mis-labeling of the device with incorrect serial number labels which may result in a patient's misdiagnosis.

Action

On 07/03/2019, the firm sent an "URGENT: Medical Device Voluntary Recall Letter" to customers via email informing them that this recall is being initiated due to the discovery of possible mis-labeling of the device packaging with an incorrect serial number label and that the firm is aware of one incident of incorrect serial number. Actions to be taken by customers are to ship back the affected capsules with serial numbers provided with the notification letter. Customer are asked to notify the Recalling Firm by email if they are experiencing inventory shortage and replacements will be shipped as soon as the firm is notified. For any the Recalling Firm directly with any concerns or questions at (408) 866-6358 Monday through Friday, 8:00 AM to 4:30 PM (Pacific Time), or at email address [email protected]

Distribution

US Nationwide distribution including in the states of FL, GA, NY, PA, and TX.

Quantity

44 units