FDA Recall Terminated

EOS X-ray Medical Diagnostic , X-ray Equipment System Digital Radiography used in general radiographic examination.

Recall: Z-2072-2012 · Initiated June 19, 2012

Recall

Recall Number
Z-2072-2012
Event Number
62621
Firm
Eos Imaging Inc
FEI Number
3008632827
Product Code
MQB
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
June 19, 2012
Posted
September 28, 2012
Terminated
September 16, 2013
Address
185 Alewife Brook Pkwy, Ste 410, Cambridge, MA, 02138-1104

Description

EOS X-ray Medical Diagnostic , X-ray Equipment System Digital Radiography used in general radiographic examination.

Reason

It was discovered of several generator failures of the EOS System due to failure of its anode controller board that drives the rotation of the X-ray tube anode.

Action

EOS Imaging notified users by letter on 6/19/12 and that an on site correction would be performed by an EOS Imaging Field Service engineer. EOS indicated in its notification that although this specific defect has not been observed in EOS systems installed in the U.S, they will replace the anode driver boards with new boards equipped with high current capacitors. Similarly, the generator firmware will be upgraded with the software patch that improves the cooling fans' controls. This will be replaced at no charge to the customers, during regularly scheduled preventative maintenance of the user's EOS system. This should be completed by the end of calendar year 2012. Your proposed CAP appears to adequately address the issues you raised in your notification, and is hereby approved. Your proposed customer notification letter as revised, appears to adequately provide instructions with respect to the use of the product pending the correction of the defect and it appears to provide a clear evaluation in nontechnical terms of the potential hazards related to the defect. Your proposed notification letter is hereby approved for distribution to all affected parties. Further questions please call 678-564-5400.

Distribution

Nationwide distribution

Quantity

12 EOS Systems installed in the US