FDA Recall Terminated

Gastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Esophagus, Stomach and Duodenum. The instrument is introduced via the mouth when indications consistent with the need for procedure are observed in adult and pediatric patient populations.

Recall: Z-2071-2016 · Initiated April 6, 2016

Recall

Recall Number
Z-2071-2016
Event Number
73894
Firm
Pentax Medical Company
FEI Number
1000080301
Product Code
FDS
Status
Terminated
Root Cause
Device Design
Initiated
April 6, 2016
Posted
June 28, 2016
Terminated
December 21, 2017
Address
3 Paragon Dr, Montvale, NJ, 07645-1782

Description

Gastroscope Intended to provide optical visualization of and therapeutic access to the Upper Gastrointestinal Tract. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Esophagus, Stomach and Duodenum. The instrument is introduced via the mouth when indications consistent with the need for procedure are observed in adult and pediatric patient populations.

Reason

Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.

Action

Pentax sent an Urgent Field Correction letter dated April 6, 2016 to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were reminded to carefully check that all accessories are intact, that no parts have fallen off and become lodged within the endoscope's instrument/suction channel and to ensure that any therapeutic devices passed through the instrument channel are accounted for after use. If the instrument/suction channel does become blocked or clogged due to the accumulation of debris, an accessory that cannot be removed, or other cause, do NOT attempt to correct the blockage or continue to use the endoscope. Customers were instructed to contact their local PENTAX Medical Service facility to have the endoscope repaired. Customers with questions were instructed to call 800-431-5880, ext 2064.

Distribution

Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam

Quantity

9,669 units