FDA Recall Terminated

Portex Tracheal Tube, Oral/Nasal, Murphy Eye, Clear, 3.5 mm Product Code: 100/127/035. Intended for oral and/or nasal intubation for airway management.

Recall: Z-2070-2009 · Initiated August 25, 2009

Recall

Recall Number
Z-2070-2009
Event Number
53075
Firm
Smiths Medical Hythe Hythe, Kent United Kingdom
FEI Number
3002807908
Product Code
BTR
Status
Terminated
Root Cause
Process control
Initiated
August 25, 2009
Posted
September 18, 2009
Terminated
November 16, 2012

Description

Portex Tracheal Tube, Oral/Nasal, Murphy Eye, Clear, 3.5 mm Product Code: 100/127/035. Intended for oral and/or nasal intubation for airway management.

Reason

Pediatric Trachel Tube manufactured with internal diameter smaller than indicated in the label may result in inability to remove secretions and cause partial or complete obstruction of the airway

Action

Smiths Medical issued an "Urgent Field Safety Notice" dated August 24, 2009 via UPS Second Day Mail including a Guidance Chart. Return of the enclosed Confirmation Form was also requested by fax to 1-603-355-8157. Smiths Medical issued a revised recall letter dated September 9, 2009. Customers are requested to return unused affected product to the firm. For further information, contact Smiths Medical ASD, Inc. by phone at 1-800-348-6064, prompt 4 extension 2273.

Distribution

Worldwide Distribution -- United States, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Borneo, Brazil, Brunei, Bulgaria, Canada, Chile, China, C¿¿te d'Ivoire, Czech Republic, Denmark, East Timor, Egypt, Estonia, Finland, France, Georgia, Germany, Gibraltar, Greece, Guatemala, Vatican City, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Malaysia, Maldives, Malta, Mexico, New Zealand, Netherlands, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom and Uruguay.

Quantity

24,582 (US) 15,620 (International)