Portex Tracheal Tube, Oral/Nasal, Murphy Eye, Siliconized, 3.0 mm Product Code: 100/141/030. Intended for oral and/or nasal intubation for airway management.
Recall
- Recall Number
- Z-2066-2009
- Event Number
- 53075
- Firm
- Smiths Medical Hythe Hythe, Kent United Kingdom
- FEI Number
- 3002807908
- Product Code
- BTR
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- August 25, 2009
- Posted
- September 18, 2009
- Terminated
- November 16, 2012
Description
Portex Tracheal Tube, Oral/Nasal, Murphy Eye, Siliconized, 3.0 mm Product Code: 100/141/030. Intended for oral and/or nasal intubation for airway management.
Pediatric Trachel Tube manufactured with internal diameter smaller than indicated in the label may result in inability to remove secretions and cause partial or complete obstruction of the airway
Smiths Medical issued an "Urgent Field Safety Notice" dated August 24, 2009 via UPS Second Day Mail including a Guidance Chart. Return of the enclosed Confirmation Form was also requested by fax to 1-603-355-8157. Smiths Medical issued a revised recall letter dated September 9, 2009. Customers are requested to return unused affected product to the firm. For further information, contact Smiths Medical ASD, Inc. by phone at 1-800-348-6064, prompt 4 extension 2273.
Worldwide Distribution -- United States, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Borneo, Brazil, Brunei, Bulgaria, Canada, Chile, China, C¿¿te d'Ivoire, Czech Republic, Denmark, East Timor, Egypt, Estonia, Finland, France, Georgia, Germany, Gibraltar, Greece, Guatemala, Vatican City, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Malaysia, Maldives, Malta, Mexico, New Zealand, Netherlands, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom and Uruguay.
95,968 (US) 496,280 (International)