FDA Recall Open, Classified

LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122

Recall: Z-2065-2026 · Initiated April 13, 2026

Recall

Recall Number
Z-2065-2026
Event Number
98418
Firm
Physio-Control, Inc.
FEI Number
3015876
Product Code
MKJ
Status
Open, Classified
Root Cause
Process control
Initiated
April 13, 2026
Posted
May 4, 2026
Address
11811 Willows Rd Ne, Redmond, WA, 98052-2003

Description

LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122

Reason

Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.

Action

On 04/13/2026, recall notices were sent to customers who were asked to do the following: 1) Return the business reply form by email to [email protected] to confirm receipt of this notification. 2) Upon receipt of the completed business reply form, firm will contact you to arrange for corrective action on the device if service has not been performed. 3) Continue to follow the Operating Instructions and perform the daily checks and user tests. These tests can be found in the chapter "Maintaining the Equipment" subsection "General Maintenance and Testing". 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. Please ensure this letter is kept with the affected device until the correction has been completed. 5) Inform firm if any of the subject devices have been distributed to other organizations. If so, provide contact details, so the firm can inform the recipients. If you have any questions or concerns, please contact the firm's Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday - Friday or by email at [email protected]

Distribution

US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA, VA, NY, MA, CO, PA, OH, SC, OK, MN, WV, NJ, WI.

Quantity

206