HeartStart MRx Processor Board PCA Replacement Kits 453563478461, a component of the Philips HeartStart MRx Monitorl/Defibrillator - Product Usage: is intended for use in hospital and prehospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support or defibrillation.

Recall: Z-2064-2020 · Initiated April 10, 2020

Recall

Recall Number
Z-2064-2020
Event Number
85462
Firm
Philips North America, LLC
FEI Number
1218950
Product Code
MKJ
Status
Terminated
Root Cause
Process control
Initiated
April 10, 2020
Terminated
February 10, 2021
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

HeartStart MRx Processor Board PCA Replacement Kits 453563478461, a component of the Philips HeartStart MRx Monitorl/Defibrillator - Product Usage: is intended for use in hospital and prehospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support or defibrillation.

Reason

Damaged ESD bags used for storing components in Processor PCA Replacement Kits

Action

Philips Healthcare ssued Medical Device Correction letter dated April 10, 2020 via USP stating reason for correction, health risk and action to take: If the Processor Board PCA Replacement Kit has not yet been installed in a Heart5tart MRx Monitor/Defibrillator, quarantine the kit until Philips can replace it. If the PCA has been installed in a HeartStart MRx Monitor/Defibrillator, continue to monitor the status of the device as recommended in the Instructions for Use, including performing Operational Checks routinely as recommended in the Instructions for Use. Verify your MRx does not show a red X nor is it emitting a periodic audio chirp as this could be an indicator a failure has been detected. When the device is turned on, the device will display INOP messages for any failures detected. If a failure has been detected, take the device out of use and contact your local Philips representative. If an affected Processor Board PCA Replacement Kit 453563478461 is identified, please contact your local Phillips representative. Once the customer has contacted their local Philips representative with their HeartStart MRx Monitor/Defibrillator serial number and also confirmed the affected Processor Board PCA serial number, Philips will provide a replacement Processor Board PCA 453563478461 Replacement Kit free of charge. Further information or support concerning this issue, contact your local Philips representative or call us at 1-800-722-9377.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of OK, NY and the countries of Libya, Netherlands, Canada, South Africa, United Kingdom.

Quantity

8 units (2 US and 6 OUS)