FDA Recall Terminated

MOSAIQ Oncology Information System MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.

Recall: Z-2063-2015 · Initiated July 1, 2015

Recall

Recall Number
Z-2063-2015
Event Number
71573
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
July 1, 2015
Posted
July 14, 2015
Terminated
September 15, 2021
Address
400 Perimeter Center Ter NE, Ste 50, Atlanta, GA, 30346-1227

Description

MOSAIQ Oncology Information System MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.

Reason

A problem exists in MOSAIQ resulting in the incorrect field size being sent to the treatment machine for stereotactic plans using cones.

Action

Elekta sent an Urgent Important Field Safety Notice 371-01-MSQ-003 on July 1, 2015, to all affected customers. The notice informed users of the specific product and version numbers affected and that the problem only exists if all the following criteria is met - (1) Running Mosaiq 2.41; (2) with at least one linac equipped with an Agility 160 MLC head (Versa HD, Infinity, Axesse (EOS) or Synergy); and (3) who are treating with Stereotactic Cones. The notice also advises users of the workaround to avoid the problem. The customers were also instructed to return the Important Field Safety Notice Acknowledgment form. For questions regarding this recall call 770-670-2422.

Distribution

Worldwide Distribution - US including AZ, CA, GA, ID, IL, MO, NJ, NY, NC, OR, SC, TN, WV, and Internationally to Austria, Brazil, Bulgaria, Canada, Colombia, Germany, Israel, Italy, Japan, Mexico, Russia, Taiwan, and Venezuela.

Quantity

74 Total (45 US)