FDA Recall Terminated

*** 1) Product labeled in part: 19734-28; LIFESHIELD; LIFESHIELD LATEX-FREE PLUM SET, MICRODRIP,PP SOLUSET,FILTER, 3 CLAVE PORTS, OPTION-LOK, NON-DEHP; NOMINAL LENGTH: 124 IN (315 IN); PRIMING VOLUME: 14 ML; *** 2) Product labeled in part: 19732-28; LIFESHIELD; LIFESHIELD LATEX-FREE PLUM SET,MICRODRIP,PP CHECK VALVE AND CLAVE OPTION-LOK, NON-DEHP; NOMINAL LENGTH:104 IN (264 CM); PRIMING VOLUME: 13 ML; Product Usage: The CLAVE of the secondary port is used as an easy secondary access that allows concurrent and piggyback mode in infusion. It is normally used with a syringe or a secondary IV infusion set.

Recall: Z-2061-2016 · Initiated September 9, 2013

Recall

Recall Number
Z-2061-2016
Event Number
66372
Firm
Hospira Inc.
FEI Number
3013319212
Product Code
FPA
Status
Terminated
Root Cause
Device Design
Initiated
September 9, 2013
Terminated
December 6, 2016
Address
275 N Field Dr, Lake Forest, IL, 60045-2579

Description

*** 1) Product labeled in part: 19734-28; LIFESHIELD; LIFESHIELD LATEX-FREE PLUM SET, MICRODRIP,PP SOLUSET,FILTER, 3 CLAVE PORTS, OPTION-LOK, NON-DEHP; NOMINAL LENGTH: 124 IN (315 IN); PRIMING VOLUME: 14 ML; *** 2) Product labeled in part: 19732-28; LIFESHIELD; LIFESHIELD LATEX-FREE PLUM SET,MICRODRIP,PP CHECK VALVE AND CLAVE OPTION-LOK, NON-DEHP; NOMINAL LENGTH:104 IN (264 CM); PRIMING VOLUME: 13 ML; Product Usage: The CLAVE of the secondary port is used as an easy secondary access that allows concurrent and piggyback mode in infusion. It is normally used with a syringe or a secondary IV infusion set.

Reason

Intravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.

Action

Hospira sent an URGENT MEDICAL DEVICE CORECTION letters dated September 9, 2013 to customers. The letter identified the affected product, problem and actions to be taken. the letter included additional instructions on the proper handling/use of the CLAVE devices. The letters request that all incidents of separation or breakage of CLAVE from the secondary port of the cassette be reported to Hospira by calling 1-800-441-410 (Monday - Friday, 8:00 AM - 5:00 PM CST) or by e-mail at [email protected]. Customers are further requested to complete and return a response form included wih the URGENT MEDICAL DEVICE CORRECTION letter and to notify any customers of this advisory. Hospira will also be supplementing the instructions for CLAVE use as an interim measure until a new method for bonding the CLAVE to the secondary port of the Plum cassettes can be devised and implemented.

Distribution

US (Nationwide) Distribution

Quantity

3,920,932 units in total