FDA Recall Terminated

Renuvion/ J-Plasma Precise Open Handpiece. Catalog numbers BVX-044-BPS, BVX-044-BPP, BVX-150-BPP, BVX-150-BPS, APYX-044-BPS, APYX-150-BPP, and APYX-150-BPS - Product Usage: for electrosurgical cutting, coagulation and ablation of soft tissue during open surgical procedures.

Recall: Z-2060-2020 · Initiated April 20, 2020

Recall

Recall Number
Z-2060-2020
Event Number
85597
Firm
Bovie Medical Corporation
FEI Number
3007593903
Product Code
GEI
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
April 20, 2020
Terminated
September 13, 2022
Address
5115 Ulmerton Rd, Clearwater, FL, 33760-4004

Description

Renuvion/ J-Plasma Precise Open Handpiece. Catalog numbers BVX-044-BPS, BVX-044-BPP, BVX-150-BPP, BVX-150-BPS, APYX-044-BPS, APYX-150-BPP, and APYX-150-BPS - Product Usage: for electrosurgical cutting, coagulation and ablation of soft tissue during open surgical procedures.

Reason

Unexpected stress fractures on the shaft of the hand piece may result in fragmentation during clinical use.

Action

On April 23, 2020 Apyx Medical (Formerly Bovie Medical) sent out an URGENT Medical Device Field Safety Notice identifying the problem, under what circumstances it can occur, the actions that should be taken by the customer/user in order to prevent risk to patients/users, and the actions planned by Apyx Medical to correct the problem. PRODUCTS: All lots of the Renuvion/J-Plasma Precise Hand piece with specified Catlog #'s that are not expired, and all lots of the Renuvion/J-Plasma Precise Open Handpiece with specified Catlog #'s that are not expired. ACTIONS TO BE TAKEN BY CUSTOMER/USER: Follow the cleaning instructions provided in the Instructions for Use (IFU) to remove coagulum/eschar from the blade during clinical use: For optimum performance, keep the distal end of the shaft free of debris. A damp gauze pad can be used for cleaning. Do not activate while cleaning the tip. Follow all specified cleaning instructions identified in the Field Safety Notification which are also in the product Instructions for Use (IFU's). Do Not modify or alter the tip of the device. Do Not use a cannula or a skin port that has tight interference with the outer diameter of the shaft of the device. Discontinue using the hand piece if during clinical use if a stress fracture develops and immediately report to Apyx Medical for return if any devices that exhibit this issue during clinical use. Customers should sign and return the customer response form via email indicating they have read and understand the instructions by May 1st of 2020.

Distribution

Worldwide distribution - US nationwide including in the states of Puerto Rico, and the countries of CA, JO, CO, MX, BR, CR, PA, KW, CL, MA, ES, TH, PH, IQ, LB, DO, Japan, QA, AE, PE, AU, VN, IL, Lithuania, UAE, Belgium, Denmark, Austria, Poland, Germany, Slovakia, Spain, Cyprus, Switzerland, Turkey, UK, Italy, Bulgaria, Greece, Norway, Ireland, Hungary, and Croatia.

Quantity

13,948