Terumo HX2 Temperature Management System 120V, 50/60 Hz catalog number 809810, Terumo Cardiovascular Systems Corp., Ann Arbor MI. The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly control the temperature of the patients blood and cardioplegia solution during cardiovascular bypass surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels.
Recall
- Recall Number
- Z-2060-2010
- Event Number
- 55224
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 1828100
- Product Code
- DWC
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 11, 2010
- Terminated
- April 13, 2011
- Address
- 6200 Jackson Road, Ann Arbor, MI, 48103-9586
Description
Terumo HX2 Temperature Management System 120V, 50/60 Hz catalog number 809810, Terumo Cardiovascular Systems Corp., Ann Arbor MI. The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly control the temperature of the patients blood and cardioplegia solution during cardiovascular bypass surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels.
The HX2 Temperature Management System mixing valve may fail resulting in an unusable water channel and then the machine would go into standby mode.
The firm, TERUMO Cardiovascular Systems, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated June 11, 2010, to the consignees. The letter describes the product, reason for correction, what to do in the event of failure, potential hazard, correction, affected population and action to be taken by the customers. The customers are instructed to review the correction notice, assure that all users are aware of this notice, and complete, sign and fax the attached Customer Response Form to Terumo CVS Customer Service at 800-292-6551. The Terumo CVS Customer Service Dept. will contact you upon receipt of the Customer Response Form to schedule the replacement. The firm plans to inspect and correct all units in the field within 5 months. We encourage you to contact Terumo CVS Customer Service at 1-800-521-2818, Monday-Friday, 8 AM-6 PM EST, with any questions or concerns.
Nationwide Distribution: AR, CA, CO, CT, FL, HI, IA, IL, KS, LA, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA and WI.
100