FDA Recall Open, Classified

da Vinci Xi/X Tip-Up Fenestrated Grasper, REF: 470347, VER: 12

Recall: Z-2059-2023 · Initiated June 12, 2023

Recall

Recall Number
Z-2059-2023
Event Number
92418
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
NAY
Status
Open, Classified
Root Cause
Process control
Initiated
June 12, 2023
Posted
June 30, 2023
Address
1266 Kifer Rd, Bldg 101, Sunnyvale, CA, 94086-5304

Description

da Vinci Xi/X Tip-Up Fenestrated Grasper, REF: 470347, VER: 12

Reason

Fenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or lower grips, which could lead to stainless steel fragments falling into the patient that may be either detected or undetected.

Action

On 6/12/23 Field Safety Notices were mailed and emailed to customers asking them to do the following: 1) Locate and return affected devices. 2) Notify all surgeons and users of the device about the contents of the notice. 3) Complete and return the acknowledgement form to [email protected] If you need further information or support concerning this Medical Device Notification, please contact your Clinical Sales Representative or contact Customer Service at: - North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: [email protected]. - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or [email protected] - South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) - Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) - India: +1-800-103-6952 (9 AM to 6 PM IT) - Taiwan: +0800-86-8181 (9 AM to 6 PM CT)

Distribution

Worldwide - US Nationwide distribution in the states of PR, FL, MN, PA, AZ, OH, IL, NC, GA, KS, CA, HI, MD, OR, SC, AK, NY, NE, IA, WA, OK, ND, UT, MS, AR, WI, MI, TX, TN, IN, NJ, VA, SD, CO, KY, AL, MO, LA, DE, MA, MT, WV, ID, CT, NV, NH, ME, WY, DC, RI, NM, VT and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Monaco, Netherlands, Norway, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.

Quantity

46,033