Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072

Recall: Z-2058-2019 · Initiated May 29, 2019

Recall

Recall Number
Z-2058-2019
Event Number
83026
Firm
Zyno Medical LLC
FEI Number
3006575795
Product Code
FPA
Status
Terminated
Root Cause
Process control
Initiated
May 29, 2019
Terminated
June 22, 2020
Address
177 Pine St, Natick, MA, 01760-1331

Description

Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072

Reason

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

Action

On May 29, 2019 Zyno issued letter to the Distributor advising of the problem and health risk and action to take: remove all recalled products with codes, cease use and immediately and return them to Zyno Medical for replacement from unaffected lots. Call or email for instructions on the return of products: (508) 907-7772 (Monday to Friday, 9:30 a.m. to 5 p.m. EST) or [email protected] to acknowledge receiving this recall notice. Zyno issued a follow-up recall letter on 6/10/19 via email to the Distributor to issue to the user/customer level.

Distribution

MA

Quantity

354 cases (50/case)=17700