TRACKMASTER, TREADMILL GE T2100-ST2 220V, MODEL NUMBER 317-07927GE
Recall
- Recall Number
- Z-2057-2023
- Event Number
- 92382
- Firm
- Full Vision Inc
- FEI Number
- 3001040337
- Product Code
- IOL
- Status
- Open, Classified
- Root Cause
- Other
- Initiated
- May 12, 2023
- Posted
- June 30, 2023
- Address
- 3017 Full Vision Dr, Newton, KS, 67114-9750
Description
TRACKMASTER, TREADMILL GE T2100-ST2 220V, MODEL NUMBER 317-07927GE
Treadmill malfunction: drive PCB can misread the signals and stop or reverse and then accelerate forward.
Trackmaster issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/12/2023 by letter. The notice explained the problem and the risk during use, and requested the following: Immediate Actions to be taken by Customer/User: These steps are the recommendation of the Manufacturer to prevent injury and allow use of the device. Step 1: Instruct patient to straddle the belt and hold on to the handrails during initialization of the treadmill before starting the test Step 2: Send start/run command to the treadmill with patient still straddling the belt, wait 30 seconds to ensure there is no uncommanded motion Step 3: Instruct patient to carefully step on the belt while holding on to the handrails, begin walking and continue with the test Step 4: Patients unable to perform steps 1-3 must not use the treadmill Step 5: Print Appendix B and attach to the treadmill until corrections are made and ensure it is visible to the clinician and user
USA (Nationwide), Australia, Bahrain, Canada, Chile, Colombia, Costa Rica, Dominican Republic, FINLAND, Guatemala, HONG KONG, India, Indonesia, Israel, Mexico, NEW ZEALAND, Nicaragua, Panama, PARAQUAY, Peru, Philippines, Saudi Arabia, Singapore, SOUTH KOREA, Thailand, Ukraine, United Arab Emirates, United Kingdom, Uruguay
1416 units